10 Monitor Spectrum Measurement
Save
Key path
SCPI Only
Remote Command
:MMEMory:COPY:DEVice <source_string>,<dest_string>
Notes The strings must be a valid logical path or a valid device keyword. If the dest_string is a device
keyword, the data is copied from the source file to the device. If the source_string is a device
keyword, the data is copied to the source file from the device.
Valid device keywords are:
SNS (smart noise source)
An error is generated if the file or device is not found.
Mass Storage Delete (Remote Command Only)
Key path
SCPI Only
Remote Command
:MMEMory:DELete <file_name>[,<directory_name>]
Notes The string must be a valid logical path.
Removes a file from the specified directory. The <file_name> parameter specifies the file name to be
removed. This command will generate an access denied” error if the file is in a restricted folder (e.g.,
C:\Windows) and the current user does not have Power User or Administrator privileges.
Initial S/W Revision Prior to A.02.00
Mass Storage Data (Remote Command Only)
Creates a file containing the specified data OR queries the data from an existing file.
Key path
SCPI Only
Remote Command
:MMEMory:DATA <file_name>, <data>
:MMEMory:DATA? <file_name>
Notes The string must be a valid logical path.
The command form is MMEMory:DATA <file_name>,<data>. It loads <data> into the file <file_
name>. <data> is in 488.2 block format. <file_name> is string data.
The query form is MMEMory:DATA? <file_name> with the response being the associated <data> in
block format.
Initial S/W Revision Prior to A.02.00
Mass Storage Make Directory (Remote Command Only)
Key path
SCPI Only
Remote Command
:MMEMory:MDIRectory <directory_name>
Notes The string must be a valid logical path.
Creates a new directory. The <directory_name> parameter specifies the name to be created.
858 EMI Receiver Mode Reference

Table of Contents

Other manuals for Keysight PXA X Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight PXA X Series and is the answer not in the manual?

Keysight PXA X Series Specifications

General IconGeneral
BrandKeysight
ModelPXA X Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals