EasyManua.ls Logo

Keysight PXA X Series - Trace Display (Remote Command Only)

Keysight PXA X Series
1138 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
10 Monitor Spectrum Measurement
Trace/ Detector
Remote Command Notes When TRACE1, the latest contents of trace 1 will be returned.
When TRACE2, the latest contents of trace 2 will be returned.
When TRACE3, the latest contents of trace 3 will be returned.
When SINGle, the current power values of all activated meters are returned separated by commas in
the order meters number.
Backwards
Compatibility SCPI
:TRACe[:DATA]?
Backwards Compatibility
SCPI Notes
This command is included for ESU compatibility. The SCPI is Meas Local and Context Sensitive.
Initial S/W Revision A.13.00
Trace Display (Remote Command Only)
This SCPI sets the state of the trace’s Update state.
Remote Command
DISPlay[:WINDow]:TRACe[1]|2|3[:STATe] ON|OFF|1|0
DISPlay[:WINDow]:TRACe[1]|2|3[:STATe]?
Example DISP:TRAC OFF
DISP:TRAC?
Notes This command is included for ESU compatibility. The suffix [1]|2 and [1]|2|3 are irrelevant to the
measurement, they are added solely for remote language compatibility. This SCPI alias to :TRACe[1]
|2|3:MONitor:UPDate ON|OFF|1|0.
Backwards
Compatibility SCPI
DISPlay:WINDow[1]|2:TRACe[1]|2|3[:STATe]
Initial S/W Revision A.13.00
EMI Receiver Mode Reference 873

Table of Contents

Other manuals for Keysight PXA X Series

Related product manuals