3 Programming the Analyzer
STATus Subsystem
Max 32767
Status Bits/OPC
dependencies
Sequential command
Initial S/W Revision Prior to A.02.00
Questionable Calibration Positive Transition
This command determines which bits in the Questionable Calibration Condition register will set the
corresponding bit in the Questionable Calibration Event register when the condition register bit has a
positive transition (0 to 1). The variable <integer> is the sum of the decimal values of the bits that you want
to enable.
Mode All
Remote Command
:STATus:QUEStionable:CALibration:PTRansition <integer>
:STATus:QUEStionable:CALibration:PTRansition?
Example STAT:QUES:CAL:PTR 16384 Alignment is required.
Preset 32767
Min 0
Max 32767
Status Bits/OPC
dependencies
Sequential command
Initial S/W Revision Prior to A.02.00
Questionable Calibration Skipped Register
"Questionable Calibration Skipped Condition " on page 118
"Questionable Calibration Skipped Enable " on page 119
"Questionable Calibration Skipped Event Query " on page 119
"Questionable Calibration Skipped Negative Transition " on page 120
"Questionable Calibration Skipped Positive Transition " on page 120
Questionable Calibration Skipped Condition
This query returns the decimal value of the sum of the bits in the Questionable Calibration Skipped
Condition register.
The data in this register is continuously updated and reflects the current conditions.
Mode All
Remote Command
:STATus:QUEStionable:CALibration:SKIPped:CONDition?
118 EMI Receiver Mode Reference