8 Frequency Scan Measurement
Meas Setup
Dependencies This key is grayed out when a scan sequence is running. If the grayed out key is pressed, an advisory
message is generated. If the equivalent SCPI command is sent, this same message is generated as
part of a “–221, Settings conflict” warning.
Preset CURRent
State Saved Saved in instrument state.
Range Curr Signal|All Signals|Marked Signals
Initial S/W Revision A.07.00
Remote Command
:CONFigure:EMI:SLISt CURRent|MARKed|ALL
Example CONF:EMI:SLIS MARK Perform remeasure on all marked signal.
Notes
CURRent aliases to :FSC:SEQ:REM CURR; :FSC:SEQ REMMARKed aliases to :FSC:SEQ:REM MARK;
:FSC:SEQ REMALL aliases to :FSC:SEQ:REM ALL; :FSC:SEQ REM
Dependencies This key is grayed out when a scan sequence is running. If the grayed out key is pressed, an advisory
message is generated. If the equivalent SCPI command is sent, this same message is generated as
part of a “–221, Settings conflict” warning.
Initial S/W Revision A.07.00
Detectors (Measure)
This key opens a menu that allows you to configure the detectors to be used for the Final Measurement
and Measure at Marker reading. The same menu can be accessed from the Detectors key under the
Marker Function menu. Any of the instruments detectors can be used for each of the three detectors, or
any of the three can be turned off. The dwell time for each detector is also settable.
When performing a Final Measurement or Measure at Marker, the dwell time settings that you select will
depend on the characteristics of the emission you are measuring. The default dwell time (200 ms) should
work well for typical EUT emissions, but sometimes you will encounter emissions for which the defaults are
not optimal. This is especially the case for emissions that vary slowly over time or have a slow repetition
rate. By lengthening the dwell times you can increase the likelihood of accurately measuring these low
repetition rate signals.
When a Final Measurement or Measure at Marker is activated, the receiver makes a zero span
measurement for each of the (up to) three detectors selected, using the Dwell Time set for each detector.
If the signals repetition period is greater than 200 ms (the default setting), the dwell time should be
increased to capture at least two and preferably more repetitions of the signal. Additionally, if you do not
need or do not wish to use a detector to make a measurement, that specific detector may be turned off.
Key Path
Meas Setup
Initial S/W Revision A.07.00
476 EMI Receiver Mode Reference

Table of Contents

Other manuals for Keysight PXA X Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight PXA X Series and is the answer not in the manual?

Keysight PXA X Series Specifications

General IconGeneral
BrandKeysight
ModelPXA X Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals