12 Disturbance Analyzer Measurement
Auto Couple
Auto Couple
The Auto Couple feature provides a quick and convenient way to automatically couple multiple instrument
settings. This helps ensure accurate measurements and optimum dynamic range. When the Auto Couple
feature is activated, either from the front panel or remotely, all parameters of the current measurement
that have an Auto/Manual mode are set to Auto mode and all measurement settings dependent on (or
coupled to) the Auto/Man parameters are automatically adjusted for optimal performance.
For the Disturbance Analyzer measurement, these parameters are:
Reference Level
Attenuation
Step Frequency
RBW
Key Path
Front-panel key