6 System Functions
System
Remote Command
:SYSTem:PON:ETIMe?
Example :SYST:PON:ETIM?
Notes Query Only
Initial S/W Revision Prior to A.02.00
Advanced
Accesses advanced diagnostic capabilities performed in the factory or under instructions from repair
procedures. This menu key is only visible when the logged-in user is “saservice”.
This feature is not available if option SF1 is installed.
The first access to the Advanced Diagnostic Menu after invoking the instrument application will require an
authentication, which is to enter the Service Code. Subsequent accesses to the Advanced Diagnostic
Menu are unimpeded. The Authentication dialog looks like:
“OK” is the default key thus the Enter key is used to complete the entry. If invalid Service Code is entered
authentication is not granted and you are provided the following dialog:
Key Path
System, Diagnostics
Notes Password is required to access this menu.
Initial S/W Revision Prior to A.02.00
Agilent Converged PSA
IP
Address
SYSTem:COMMunicate:LAN:ADDRess <string>
SYSTem:COMMunicate:LAN:ADDRess?
:SYSTem:COMMunicate:LAN[:SELF]:IP <string>
:SYSTem:COMMunicate:LAN[:SELF]:IP?
Gateway SYSTem:COMMunicate:LAN:DGATeway <string> :SYSTem:COMMunicate:LAN[:SELF]:GATEway <string