8 Frequency Scan Measurement
Meas Setup
Dependencies This key is grayed out before Marker Zoom is performed. If the grayed out key is pressed, an advisory
message is generated. If the equivalent SCPI command is sent, this same message is generated as
part of a “–200, Execution error” warning.
This key is grayed out when a scan sequence is running. If the grayed out key is pressed, an advisory
message is generated. If the equivalent SCPI command is sent, this same message is generated as
part of a “–221, Settings conflict” warning.
Initial S/W Revision A.07.00
Sort Signals
Specifies how the elements in the signal list will be sorted either based on the frequency, detector
amplitude, detector vs. limit delta or time in ascending or descending order. During sorting, an advisory
message appears at the message bar, Sorting signal list by <sorting type selected>…”, for example:
Sorting signal list by Frequency…
Sorting signal list by Det 1 Amplitude
Sorting signal list by Det 3 vs Limit Delta
Sorting signal list by Det 3 vs TimeStamp…
Key Path
Meas Setup, Signal List, Sort Signals
Remote Command
:CALCulate:FSCan:SLISt:SORT:TYPE FREQuency | DAMPlitude | DLDelta |
TIME
:CALCulate:FSCan:SLISt:SORT:TYPE?
Example CALC:FSC:SLIS:SORT:TYPE FREQ
CALC:FSC:SLIS:SORT:TYPE?
Dependencies This key is grayed out when a scan sequence is running. If the grayed out key is pressed, an advisory
message is generated. If the equivalent SCPI command is sent, this same message is generated as
part of a “–221, Settings conflict” warning.
Couplings Selecting a sorting mode (pressing the key or sending the equivalent SCPI command) will turn on the
Auto Sort state, even if that sorting mode was already selected.
Preset FREQuency
State Saved Saved in instrument state.
Range By Freq|By Detector Amptd|By ( Detector|By Time
Initial S/W Revision A.07.00
By Freq
Specifies that the elements in the signal list will be sorted based on the frequency.
Key Path
Meas Setup, Signal List, Signal List, Sort Signals
Example CALC:FSC:SLIS:SORT:TYPE FREQ
Initial S/W Revision A.07.00
448 EMI Receiver Mode Reference

Table of Contents

Other manuals for Keysight PXA X Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight PXA X Series and is the answer not in the manual?

Keysight PXA X Series Specifications

General IconGeneral
BrandKeysight
ModelPXA X Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals