10 Monitor Spectrum Measurement
Meas Setup
It is possible to change this setting after a limit line has been entered. When changing from On to Off or
vice-versa, the amplitude values in the limit line table change so that the limit line remains in the same
position for the current reference level settings of the instrument.
Key Path
Meas Setup, Limits, Properties, Fixed/Relative
Remote Command
:CALCulate:LLINe[1]|2|...|6:AMPLitude:CMODe:RELative ON | OFF | 1 | 0
:CALCulate:LLINe[1]|2|...|6:AMPLitude:CMODe:RELative?
Example CALC:LLIN:AMPL:CMOD:REL ON Makes limit line 1 relative to the reference level amplitude.
Couplings Pressing this button makes Reference level the active function.
Preset Off, not affected by Mode Preset, preset by Restore Mode Defaults.
State Saved Saved in instrument state
Range On|Off
Initial S/W Revision A.13.00
Description
Provides a description of up to 60 characters by which you can easily identify the limit. The descriptions will
be stored in the exported file and can be displayed in the active function area by selecting them as the
active function, if desired to be in a screen dump.
Key Path
Meas Setup, Limits, Properties
Remote Command
:CALCulate:LLINe[1]|2|...|6:DESCription “string
:CALCulate:LLINe[1]|2|...|6:DESCription?
Example CALC:LLIN:DESC "European Emissions"
Preset (null String), not affected by Mode Preset, preset by Restore Mode Defaults.
State Saved Saved in instrument state
Backwards
Compatibility SCPI
:CALCulate[1]|2:LIMit[1]|2|...|6:NAME
Backwards Compatibility
SCPI Notes
The suffix [1]|2 is irrelevant to the measurement, they are added solely for remote language
compatibility.
Initial S/W Revision A.13.00
Comment
Sets an ASCII comment field, which will be stored in an exported file. The comment can be displayed in the
active function area by selecting it as the active function, if desired to be in a screen dump. The Limits .csv
file supports this field.
Key Path
Meas Setup, Limits, Properties
Remote Command
:CALCulate:LLINe[1]|2|...|6:COMMent “string”
:CALCulate:LLINe[1]|2|...|6:COMMent ?
EMI Receiver Mode Reference 795

Table of Contents

Other manuals for Keysight PXA X Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight PXA X Series and is the answer not in the manual?

Keysight PXA X Series Specifications

General IconGeneral
BrandKeysight
ModelPXA X Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals