6 System Functions
System
Example :CAL:RFPS:SCH:STAT OFF
Notes For model N9038A only.
Preset This is unaffected by Preset but is set to OFF on a Restore System Defaults->Align.
State Saved No
Initial S/W Revision A.08.00
I/O Config
Activates a menu for identifying and changing the I/O configuration for remote control.
Key Path
System
Initial S/W Revision Prior to A.02.00
GPIB
Activates a menu for configuring the GPIB I/O port.
Key Path
System, I/O Config
Initial S/W Revision A.02.00
GPIB Address
Select the GPIB remote address.
Key Path
System, I/O Config, GPIB
Mode All
Remote Command
:SYSTem:COMMunicate:GPIB[1][:SELF]:ADDRess <integer>
:SYSTem:COMMunicate:GPIB[1][:SELF]:ADDRess?
Example :SYST:COMM:GPIB:ADDR 17
Notes Changing the Address on the GPIB port requires all further communication to use the new address.
Preset This is unaffected by Preset but is set to 18 on a Restore System Defaults->Misc”
State Saved No
Range 0 to 30
Min 0
Max 30
Initial S/W Revision Prior to A.02.00
332 EMI Receiver Mode Reference

Table of Contents

Other manuals for Keysight PXA X Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight PXA X Series and is the answer not in the manual?

Keysight PXA X Series Specifications

General IconGeneral
BrandKeysight
ModelPXA X Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals