3 Programming the Analyzer
STATus Subsystem
The register requires that the associated PTR or NTR filters be set before a condition register bit can set a
bit in the event register. The data in this register is latched until it is queried. Once queried, the register is
cleared
Mode All
Remote Command
:STATus:QUEStionable:TEMPerature[:EVENt]?
Example STAT:QUES:TEMP?
Preset 0
Status Bits/OPC
dependencies
Sequential command
Initial S/W Revision Prior to A.02.00
Questionable Temperature Negative Transition
This command determines which bits in the Questionable Temperature Condition register will set the
corresponding bit in the Questionable Temperature Event register when the condition register bit has a
negative transition (1 to 0). The variable <integer> is the sum of the decimal values of the bits that you want
to enable.
Mode All
Remote Command
:STATus:QUEStionable:TEMPerature:NTRansition <integer>
:STATus:QUEStionable:TEMPerature:NTRansition?
Example STAT:QUES:TEMP:NTR 1 Reference Oscillator Oven not cold will be reported to the Temperature
Summary of the Status Questionable register.
Preset 0
Min 0
Max 32767
Status Bits/OPC
dependencies
Sequential command
Initial S/W Revision Prior to A.02.00
Questionable Temperature Positive Transition
This command determines which bits in the Questionable Temperature Condition register will set the
corresponding bit in the Questionable Temperature Event register when the condition register bit has a
positive transition (0 to 1). The variable <integer> is the sum of the decimal values of the bits that you want
to enable.
Mode All
Remote Command
:STATus:QUEStionable:TEMPerature:PTRansition <integer>
:STATus:QUEStionable:TEMPerature:PTRansition?
Example STAT:QUES:TEMP:PTR 1 Reference Oscillator Oven became cold will be reported to the
138 EMI Receiver Mode Reference