10 Monitor Spectrum Measurement
Meas Setup
Upper
Marks duplicate signals with higher amplitude.
Key Path
Meas Setup, Signal List, Mark Signals, Mark Duplicates
Remote Command
:CALCulate:SLISt:MARK:DUPLicates:UPPer
Example CALC:SLIS:MARK:DUPL:UPP
Notes If the signal list contains no data, an error is generated and this function cannot be performed.
Initial S/W Revision A.13.00
Lower
Marks duplicate signals with lower amplitude.
Key Path
Meas Setup, Signal List, Mark Signals, Mark Duplicates
Remote Command
:CALCulate:SLISt:MARK:DUPLicates:LOWer
Example CALC:SLIS:MARK:DUPL:LOW
Notes If the signal list contains no data, an error is generated and this function cannot be performed.
Initial S/W Revision A.13.00
Clear Mark
Clears mark from the selected signal.
Key Path
Meas Setup, Signal List, Mark Signals
Remote Command
:CALCulate:SLISt:MARK:CLEar:SIGNal <integer>
Example CALC:SLIS:MARK:CLE:SIGN 11 Clear mark from signal #11.
Notes If the signal list contains no data, an error is generated and this function cannot be performed.
If the selected signal is not in the list, an error is generated and this function cannot be performed.
Initial S/W Revision A.13.00
Mark All
Marks all signals in the signal list.
Key Path
Meas Setup, Signal List, Mark Signals
Remote Command
:CALCulate:SLISt:MARK:ALL
Example CALC:SLIS:MARK:ALL
Notes If the signal list contains no data, an error is generated and this function cannot be performed.
Initial S/W Revision A.13.00