6 System Functions
System
Key Path
System, Alignments, Auto Align, Alert
Mode All
Example :CAL:AUTO:ALER WEEK
Readback Text 7 days
Status Bits/OPC
dependencies
Bit 14 is set in the Status Questionable Calibration register.
Initial S/W Revision Prior to A.02.00
None
With Auto Align Alert set to None the instrument will not signal an alert. This is provided for rare occasions
where you are making a long measurement which cannot tolerate Auto Align interruptions, and must have
the ability to capture a screen image at the end of the measurement without an alert posted to the display.
Agilent does not recommends using this selection in any other circumstances, because of the risk of
accuracy performance drifting well beyond expected levels without the operator being informed.
For front panel operation, confirmation is required to transition into this setting of Alert. The confirmation
dialog is:
No confirmation is required when Alert is configured through a remote command.
Key Path
System, Alignments, Auto Align, Alert
Mode All
Example :CAL:AUTO:ALER NONE
Initial S/W Revision Prior to A.02.00
Execute Expired Alignments (Remote Command Only)
Alignments can be expired in the situation where Auto Align is in the state of Partial or Off. This feature runs
the alignments that have expired. This is different than performing an Align All, Now operation. Align All,
Now performs an alignment of all subsystems regardless of whether they are needed or not, with Execute
Expired Alignments, only the individual subsystems that have become due are aligned.
296 EMI Receiver Mode Reference

Table of Contents

Other manuals for Keysight PXA X Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight PXA X Series and is the answer not in the manual?

Keysight PXA X Series Specifications

General IconGeneral
BrandKeysight
ModelPXA X Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals