EasyManuals Logo

Keysight PXA X Series User Manual

Keysight PXA X Series
1138 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #443 background imageLoading...
Page #443 background image
8 Frequency Scan Measurement
Meas Setup
All
Marks all duplicate signals.
Key Path
Meas Setup, Signal List, Mark Signals, Mark Duplicates
Remote Command
:CALCulate:FSCan:SLISt:MARK:DUPLicates[:ALL]
Example CALC:FSC:SLIS:MARK:DUPL
Notes If the signal list contains no data, an error is generated and this function cannot be performed.
Dependencies This key is grayed out when a scan sequence is running. If the grayed out key is pressed, an advisory
message is generated. If the equivalent SCPI command is sent, this same message is generated as
part of a “–221, Settings conflict” warning.
Initial S/W Revision A.13.00
Upper
Marks duplicate signals with higher amplitudes.
Key Path
Meas Setup, Signal List, Mark Signals, Mark Duplicates
Remote Command
:CALCulate:FSCan:SLISt:MARK:DUPLicates:UPPer
Example CALC:FSC:SLIS:MARK:DUPL:UPP
Notes If the signal list contains no data, an error is generated and this function cannot be performed.
Dependencies This key is grayed out when a scan sequence is running. If the grayed out key is pressed, an advisory
message is generated. If the equivalent SCPI command is sent, this same message is generated as
part of a “–221, Settings conflict” warning.
Initial S/W Revision A.13.00
Lower
Marks duplicate signals with lower amplitudes.
Key Path
Meas Setup, Signal List, Mark Signals, Mark Duplicates
Remote Command
:CALCulate:FSCan:SLISt:MARK:DUPLicates:LOWer
Example CALC:FSC:SLIS:MARK:DUPL:LOW
Notes If the signal list contains no data, an error is generated and this function cannot be performed.
Dependencies This key is grayed out when a scan sequence is running. If the grayed out key is pressed, an advisory
message is generated. If the equivalent SCPI command is sent, this same message is generated as
part of a “–221, Settings conflict” warning.
Initial S/W Revision A.13.00
Clear Mark
Clears the mark from the selected signal.
EMI Receiver Mode Reference 443

Table of Contents

Other manuals for Keysight PXA X Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight PXA X Series and is the answer not in the manual?

Keysight PXA X Series Specifications

General IconGeneral
BrandKeysight
ModelPXA X Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals