8 Frequency Scan Measurement
Meas Setup
All
Marks all duplicate signals.
Key Path
Meas Setup, Signal List, Mark Signals, Mark Duplicates
Remote Command
:CALCulate:FSCan:SLISt:MARK:DUPLicates[:ALL]
Example CALC:FSC:SLIS:MARK:DUPL
Notes If the signal list contains no data, an error is generated and this function cannot be performed.
Dependencies This key is grayed out when a scan sequence is running. If the grayed out key is pressed, an advisory
message is generated. If the equivalent SCPI command is sent, this same message is generated as
part of a “–221, Settings conflict” warning.
Initial S/W Revision A.13.00
Upper
Marks duplicate signals with higher amplitudes.
Key Path
Meas Setup, Signal List, Mark Signals, Mark Duplicates
Remote Command
:CALCulate:FSCan:SLISt:MARK:DUPLicates:UPPer
Example CALC:FSC:SLIS:MARK:DUPL:UPP
Notes If the signal list contains no data, an error is generated and this function cannot be performed.
Dependencies This key is grayed out when a scan sequence is running. If the grayed out key is pressed, an advisory
message is generated. If the equivalent SCPI command is sent, this same message is generated as
part of a “–221, Settings conflict” warning.
Initial S/W Revision A.13.00
Lower
Marks duplicate signals with lower amplitudes.
Key Path
Meas Setup, Signal List, Mark Signals, Mark Duplicates
Remote Command
:CALCulate:FSCan:SLISt:MARK:DUPLicates:LOWer
Example CALC:FSC:SLIS:MARK:DUPL:LOW
Notes If the signal list contains no data, an error is generated and this function cannot be performed.
Dependencies This key is grayed out when a scan sequence is running. If the grayed out key is pressed, an advisory
message is generated. If the equivalent SCPI command is sent, this same message is generated as
part of a “–221, Settings conflict” warning.
Initial S/W Revision A.13.00
Clear Mark
Clears the mark from the selected signal.
EMI Receiver Mode Reference 443