3 Programming the Analyzer
STATus Subsystem
"Questionable Calibration Extended Failure Positive Transition " on page 122
Questionable Calibration Extended Failure Condition
This query returns the decimal value of the sum of the bits in the Questionable Calibration Extended Failure
Condition register.
The data in this register is continuously updated and reflects the current conditions.
Mode All
Remote Command
:STATus:QUEStionable:CALibration:EXTended:FAILure:CONDition?
Example STAT:QUES:CAL:EXT:FAIL:COND?
Preset 0
Status Bits/OPC
dependencies
Sequential command
Initial S/W Revision Prior to A.02.00
Questionable Calibration Extended Failure Enable
This command determines which bits in the Questionable Calibration Extended Failure Condition Register
will set bits in the Questionable Calibration Extended Failure Event register, which also sets bit 9 of the
Questionable Calibration Register. The variable <integer> is the sum of the decimal values of the bits you
want to enable.
Mode All
Remote Command
:STATus:QUEStionable:CALibration:EXTended:FAILure:ENABle <integer>
:STATus:QUEStionable:CALibration:EXTended:FAILure:ENABle?
Example STAT:QUES:CAL:EXT:FAIL:ENAB 1 Can be used to query if an EMI conducted alignment is needed.
Preset 32767
Min 0
Max 32767
Status Bits/OPC
dependencies
Sequential command
Initial S/W Revision Prior to A.02.00
Questionable Calibration Extended Failure Event Query
This query returns the decimal value of the sum of the bits in the Questionable Calibration Extended Failure
Event register.
The register requires that the associated PTR or NTR filters be set before a condition register bit can set a
bit in the event register. The data in this register is latched until it is queried. Once queried, the register is
cleared.
EMI Receiver Mode Reference 121