10 Monitor Spectrum Measurement
Meas Setup
Example CALC:SLIS:SORT:TYPE FREQ
Initial S/W Revision A.13.00
By Detector Amptd
Specifies that the elements in the signal list will be sorted based on the detector amplitude.
Key Path
Meas Setup, Signal List, Sort Signals
Example CALC:SLIS:SORT:TYPE DAMP
Initial S/W Revision A.13.00
By Det Amptd Type
Specifies how the signal list will be sorted based on the detector amplitude type.
Key Path
Meas Setup, Signal List, Sort Signals, By Detector Amplitude
Remote Command
:CALCulate:SLISt:SORT:DAMPlitude DET1|DET2|DET3
:CALCulate:SLISt:SORT:DAMPlitude ?
Example CALC:SLIS:SORT:DAMP DET2
CALC:SLIS:SORT:DAMP?
Dependencies Selecting the detector amplitude type to be sorted (pressing the key or sending the equivalent SCPI
command) will turn on the Auto Sort state, even if that detector amplitude type was already
selected.
Preset DET1
State Saved Saved in instrument state
Range Detector 1|Detector 2|Detector 3
Initial S/W Revision A.13.00
By Δ Detector
Specifies that the elements in the signal list will be sorted based on the detector vs limit delta.
Key Path
Meas Setup, Signal List, Sort Signals
Example CALC:SLIS:SORT:DET DLD
Initial S/W Revision A.13.00
By Δ Detector Type
Specifies how the signal list will be sorted based on the detector to limit delta.
EMI Receiver Mode Reference 787