EasyManuals Logo

Keysight PXA X Series User Manual

Keysight PXA X Series
1138 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #521 background imageLoading...
Page #521 background image
8 Frequency Scan Measurement
Quick Save
Quick Save
The Quick Save front-panel key repeats the most recent save that was performed from the Save menu,
with the following exceptions:
Register saves are not remembered as Saves for the purpose of the Quick Save function
If the current measurement does not support the last non-register save that was performed, an
informational message is generated, File type not supported for this measurement”
Quick Save repeats the last type of qualified save (that is, a save qualified by the above criteria) in the last
save directory by creating a unique filename using the Auto File Naming algorithm described below.
If Quick Save is pressed after startup and before any qualified Save has been performed, the Quick Save
function performs a Screen Image save using the current settings for Screen Image saves (current theme,
current directory), which then becomes the last save for the purpose of subsequent Quick Saves.
The Auto File Naming feature automatically generates a file name for use when saving a file. The filename
consists of a prefix and suffix separated by a dot, as is standard for the Windows® file system. A default
prefix exists for each of the available file types:
Type Default Prefix Menu
State State_ (Save/Recall)
Trace + State State_ (Save/Recall)
Screen Screen_ (Save/Recall)
Amplitude Corrections Ampcor_ (Import/Export)
Traces Trace_ (Import/Export)
Limit Lines LLine_ (Import/Export)
Measurement Result MeasR_ (Import/Export)
Capture Buffer CapBuf_ (Import/Export)
A four digit number is appended to the prefix to create a unique file name. The numbering sequence starts
at 0000 within each Mode for each file type and updates incrementally to 9999, then wraps to 0000 again.
It remembers where it was through a Mode Preset and when leaving and returning to the Mode. It is reset
by Restore Misc Defaults and Restore System Defaults and subsequent running of the instrument
application. So, for example, the first auto file name generated for State files is State_0000.state. The next
is State_0001, and so forth.
One of the key features of Auto File Name is that we guarantee that the Auto File Name will never conflict
with an existing file.The algorithm looks for the next available number. If it gets to 9999, then it looks for
holes. If it find no holes, that is no more numbers are available, it gives an error.
For example, if when we get to State_0010.state there is already a State_0010.state file in the current
directory, it advances the counter to State_0011.state to ensure that no conflict will exist (and then it
verifies that State_0011.state also does not exist in the current directory and advances again if it does, and
so forth).
EMI Receiver Mode Reference 521

Table of Contents

Other manuals for Keysight PXA X Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight PXA X Series and is the answer not in the manual?

Keysight PXA X Series Specifications

General IconGeneral
BrandKeysight
ModelPXA X Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals