8 Frequency Scan Measurement
Sweep/ Control
Remote Command
INITiate[1]:CONTinuous OFF
Example INIT:CONT 0; INIT:IMM
Dependencies This key is grayed out when a scan sequence is running. If the grayed out key is pressed, an advisory
message is generated. If the equivalent SCPI command is sent, this same message is generated as
part of a “–221, Settings conflict” warning.
Couplings Meters will do a single measurement and stop. You can start the meters by changing the setting to
continuous, or by sending the remote command INIT:IMM .
Initial S/W Revision A.07.00
Continuous (Meters)
Performs a continuous measurement at meters.
Key Path
Sweep/Control
Remote Command
INITiate[1]:CONTinuous ON
Example INIT:CONT 1
Dependencies This key is grayed out when a scan sequence is running. If the grayed out key is pressed, an advisory
message is generated. If the equivalent SCPI command is sent, this same message is generated as
part of a “–221, Settings conflict” warning.
Couplings Meters will do measurements continuously, until you change the setting to single or a scan sequence
is started.
Initial S/W Revision A.07.00
Restart Meters (Remote Command Only)
Allows you to restart the meters measurement when set to single.
Remote Command
INITiate[1]:IMMediate
INITiate[1]:RESTart
Example
INIT:IMMINIT:REST
Notes This command is remote language compatible with ESU.
Initial S/W Revision A.07.00