6 System Functions
System
Restore Align Defaults
This selection causes the Alignment system settings to be a reset to their default values. This does not
affect any Alignment data stored in the system. This level of Restore System Defaults does not affect any
other system settings, mode settings and does not cause a mode switch.
After performing this function, it may impact the auto-alignment time of the instrument until a new
alignment baseline has been established.
Confirmation is required to restore the factory default values. The confirmation dialog is:
Key Path
System, Restore System Defaults
Example :SYST:DEF ALIG
Initial S/W Revision Prior to A.02.00
Restore Misc Defaults
This selection causes miscellaneous system settings to be reset to their default values. With this reset,
you lose the GPIB address and it is reset to 18, so this should be used with caution. This level of Restore
System Defaults does not affect any other system settings, mode settings and does not cause a mode
switch. This miscellaneous group contains the rest of the settings that have not been part of the other
Restore System Defaults groups. The following table is a complete list of settings associated with this
group:
Miscellaneous Setting Default Value
Verbose SCPI Off
The SYST:PRES:TYPE MODE
Auto File Name Number 000
Save Type State
State Save To Register 1
Screen Save To SCREEN000.png
DISP:ENABle ON
Full Screen Off
342 EMI Receiver Mode Reference