3 Programming the Analyzer
STATus Subsystem
Status Bits/OPC
dependencies
Sequential command
Initial S/W Revision Prior to A.02.00
Questionable Calibration Extended Needed Register
"Questionable Calibration Extended Needed Condition " on page 123
"Questionable Calibration Extended Needed Enable " on page 123
"Questionable Calibration Extended Needed Event Query " on page 124
"Questionable Calibration Extended Needed Negative Transition " on page 124
"Questionable Calibration Extended Needed Positive Transition " on page 125
Questionable Calibration Extended Needed Condition
This query returns the decimal value of the sum of the bits in the Questionable Calibration Extended
Needed Condition register.
The data in this register is continuously updated and reflects the current conditions.
Mode All
Remote Command
:STATus:QUEStionable:CALibration:EXTended:NEEDed:CONDition?
Example STAT:QUES:CAL:EXT:NEED:COND?
Preset 0
Status Bits/OPC
dependencies
Sequential command
Initial S/W Revision Prior to A.02.00
Questionable Calibration Extended Needed Enable
This command determines which bits in the Questionable Calibration Extended Needed Condition Register
will set bits in the Questionable Calibration Extended Needed Event register, which also sets bit 14 of the
Questionable Calibration Register. The variable <integer> is the sum of the decimal values of the bits you
want to enable.
Mode All
Remote Command
:STATus:QUEStionable:CALibration:EXTended:NEEDed:ENABle <integer>
:STATus:QUEStionable:CALibration:EXTended:NEEDed:ENABle?
Example STAT:QUES:CAL:EXT:NEED:ENAB 2 Can be used to query if an EMI conducted alignment is needed.
Preset 32767
Min 0
EMI Receiver Mode Reference 123