12 Disturbance Analyzer Measurement
Meas Setup
Meas Setup
Displays the menu keys that enable you to setup the measurement.
Key Path
Front-panel key
Notes Front panel only.
Initial S/W Revision A.14.00
Select Disturbance
Specifies the selected disturbance.
Key Path
Meas Setup, Disturbance List
Notes Front panel only. When select disturbance is the active function, press UP to go up in the disturbance
list, and press DOWN to go down in the list.
Dependencies Yes
Couplings When the disturbance list is cleared, this value will reset to 0.
Preset If the disturbance list is empty, =0
If the disturbance list is not empty, = 1
State Saved Saved in instrument state
Min 0
Max Number of disturbances in disturbance list
Initial S/W Revision A.14.00
Disturbance List
This menu enables you to access the disturbance list related functions.
Key Path
Meas Setup
Initial S/W Revision A.14.00
Select Disturbance
Specifies the selected disturbance.
Key Path
Meas Setup, Disturbance List
Notes Front panel only. When select disturbance is the active function, press UP to go up in the disturbance
list, and press DOWN to go down in the list.
Dependencies Yes
Couplings When the disturbance list is cleared, this value will reset to 0.
EMI Receiver Mode Reference 1059