5 Mode Functions
Mode Setup
SENS:FSC:SLIS:COUP:MET?
Couplings Couple Meters to Marker key will be turned off when this key is turned on.
Preset ON
State Saved Saved in instrument state.
Range On|Off
Initial S/W Revision A.07.00
Couple Meters to Marker
If this is turned on, the measurement will automatically couple Meters frequency to frequency of the
selected marker.
Key Path
Marker
Remote Command
:CALCulate:FSCan:MARKer:COUPle:METer ON|OFF|1|0
:CALCulate:FSCan:MARKer:COUPle:METer?
Example CALC:FSC:MARK:COUP:MET 1
CALC:FSC:MARK:COUP:MET?
Couplings Couple Meters to Signal List key will be turned off when this key is turned on.
Preset OFF
State Saved Saved in instrument state.
Range On|Off
Backwards
Compatibility SCPI
:CALCulate[1]|2:MARKer:COUPled[:STATe] ON|OFF|1|0
:CALCulate[1]|2:MARKer:COUPled[:STATe]?
Initial S/W Revision A.13.00
Autorange Atten 10 dB Minimum
Toggles the autorange attenuation protection to have at least 10 dB always set on the input.
Key Path
Mode Setup
Remote Command
[:SENSe]:METer:ATTenuation:PROTection ON|OFF|1|0
[:SENSe]:METer:ATTenuation:PROTection?
Example MET:ATT:PROT ON
MET:ATT:PROT?
Couplings When ON, an attenuation of at least 10 dB is always set on the input when autorange is running
during meters measurement, prescan and the final measurement.
Preset OFF
State Saved Saved in instrument state
Backwards
:INPut[1]|2:ATTenuation:PROTection