11 APD Measurement
Save
Key path
SCPI Only
Remote Command
:MMEMory:DATA <file_name>, <data>
:MMEMory:DATA? <file_name>
Notes The string must be a valid logical path.
The command form is MMEMory:DATA <file_name>,<data>. It loads <data> into the file <file_
name>. <data> is in 488.2 block format. <file_name> is string data.
The query form is MMEMory:DATA? <file_name> with the response being the associated <data> in
block format.
Initial S/W Revision Prior to A.02.00
Mass Storage Make Directory (Remote Command Only)
Key path
SCPI Only
Remote Command
:MMEMory:MDIRectory <directory_name>
Notes The string must be a valid logical path.
Creates a new directory. The <directory_name> parameter specifies the name to be created.
This command will generate an access denied” error if the new directory would be in a restricted
folder (e.g., C:\Windows) and the current user does not have Power User or Administrator privileges.
Initial S/W Revision Prior to A.02.00
Mass Storage Move (Remote Command Only)
Key path
SCPI Only
Remote Command
:MMEMory:MOVE <string>,<string>[,<string>,<string>]
Notes The string must be a valid logical path.
Moves an existing file to a new file or an existing directory to a new directory.
Two forms of parameters are allowed. The first form has two parameters. In this form, the first
parameter specifies the source, and the second parameter specifies the destination.
The second form has four parameters. In this form, the first and third parameters specify the source.
The second and fourth parameters specify the directories. The first pair of parameters specifies the
source. The second pair specifies the destination. An error is generated if the source doesn’t exist or
the destination file already exists.
This command will generate an access denied” error if the destination is a restricted folder (e.g.,
C:\Windows) and the current user does not have Power User or Administrator privileges.
Initial S/W Revision Prior to A.02.00
EMI Receiver Mode Reference 999

Table of Contents

Other manuals for Keysight PXA X Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight PXA X Series and is the answer not in the manual?

Keysight PXA X Series Specifications

General IconGeneral
BrandKeysight
ModelPXA X Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals