3 Programming the Analyzer
IEEE 488.2 Common Commands
Wait-to-Continue
This command causes the instrument to wait until all overlapped commands are completed before
executing any additional commands. There is no query form for the command.
Remote Command
*WAI
Example INIT:CONT OFF; INIT;*WAI Sets the instrument to single sweep. Starts a sweep and waits for its
completion.
Status Bits/OPC
dependencies
Not global to all remote ports or front panel. *OPC only considers operation that was initiated on the
same port as the *OPC command was issued from.
Initial S/W Revision Prior to A.02.00