EasyManuals Logo

Keysight PXA X Series User Manual

Keysight PXA X Series
1138 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #836 background imageLoading...
Page #836 background image
10 Monitor Spectrum Measurement
Restart
Restart
The Restart function restarts the current sweep, or measurement, or set of averaged/held sweeps or
measurements. If you are Paused, pressing Restart does a Resume.
The Restart function is accessed in several ways:
Pressing the Restart key
Sending the remote command INIT:IMMediate
Sending the remote command INIT:RESTart
See "More Information" on page 836
Key Path
Front-panel key
Remote Command
:INITiate[:IMMediate]
:INITiate:RESTart
Example :INIT:IMM
:INIT:REST
Notes :INITiate:RESTart and :INITiate:IMMediate perform exactly the same function.
Couplings Resets average/hold count k. For the first sweep overwrites all active (update=on) traces with new
current data. For application modes, it resets other parameters as required by the measurement.
Status Bits/OPC
dependencies
This is an Overlapped command.
The STATus:OPERation register bits 0 through 8 are cleared.
The STATus:QUEStionable register bit 9 (INTegrity sum) is cleared.
The SWEEPING bit is set.
The MEASURING bit is set.
Backwards Compatibility
Notes
For Spectrum Analysis mode in ESA and PSA, the Restart hardkey and the INITiate:RESTart
command restart trace averages (displayed average count reset to 1) for a trace in Clear Write, but
did not restart Max Hold and Min Hold.
In the X-Series, the Restart hardkey and the INITiate:RESTart command restart not only Trace
Average, but MaxHold and MinHold traces as well.
For wireless comms modes in ESA and PSA, the Restart hardkey and the INITiate:RESTart command
restart every measurement, which includes all traces and numeric results. There is no change to this
operation.
Initial S/W Revision Prior to A.02.00
More Information
The Restart function first aborts the current sweep/measurement as quickly as possible. It then resets the
sweep and trigger systems, sets up the measurement and initiates a new data measurement sequence
with a new data acquisition (sweep) taken once the trigger condition is met.
If the analyzer is in the process of aligning when Restart is executed, the alignment finishes before the
restart function is performed.
836 EMI Receiver Mode Reference

Table of Contents

Other manuals for Keysight PXA X Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight PXA X Series and is the answer not in the manual?

Keysight PXA X Series Specifications

General IconGeneral
BrandKeysight
ModelPXA X Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals