6 System Functions
System
Example :LXI:IDEN ON
Preset Not part of Preset, but reset to OFF on Restore System Defaults All
State Saved No
Range On | Off
Initial S/W Revision A.12.50
System IDN Response
This key allows you to specify a response to the *IDN? query, or to return the analyzer to the Factory
response if you have changed it.
To choose the factory-set response, press the Factory key.
To specify your own response, press the User key, and enter your desired response.
Key Path
System, I/O Config
Mode All
Remote Command
:SYSTem:IDN <string>
:SYSTem:IDN?
Notes
• This affects the response given in all Modes of the Analyzer, unless the current Mode has also
specified a custom response, in which case the current Mode’s custom IDN response takes
precedence over the System’s, but only while that Mode is the current Mode..
• It survives shutdown and restart of the software and therefore survives a power cycle
• Null string as parameter restores the Factory setting
Preset This is unaffected by Preset but is set to the original factory setting on a “Restore System Defaults-
>Misc”
State Saved No
Initial S/W Revision A.06.00
Factory
This key selects the factory setting, for example:
“Agilent Technologies,N9020A,MY00012345,A.05.01”
where the fields are manufacturer, model number, serial number, firmware revision.
Key Path
System, I/O Config, IDN Response
Example :SYST:IDN "" null string, restores the factory setting
Initial S/W Revision A.06.0
338 EMI Receiver Mode Reference