3 Programming the Analyzer
STATus Subsystem
Max 32767
Status Bits/OPC
dependencies
Sequential command
Initial S/W Revision Prior to A.02.00
Questionable Calibration Extended Needed Event Query
This query returns the decimal value of the sum of the bits in the Questionable Calibration Extended
Needed Event register.
The register requires that the associated PTR or NTR filters be set before a condition register bit can set a
bit in the event register. The data in this register is latched until it is queried. Once queried, the register is
cleared.
Mode All
Remote Command
:STATus:QUEStionable:CALibration:EXTended:NEEDed[:EVENt]?
Example STAT:QUES:CAL:EXT:NEED?
Preset 0
Status Bits/OPC
dependencies
Sequential command
Initial S/W Revision Prior to A.02.00
Questionable Calibration Extended Needed Negative Transition
This command determines which bits in the Questionable Calibration Extended Needed Condition register
will set the corresponding bit in the Questionable Calibration Extended Needed Event register when the
condition register bit has a negative transition (1 to 0). The variable <integer> is the sum of the decimal
values of the bits that you want to enable.
Mode All
Remote Command
:STATus:QUEStionable:CALibration:EXTended:NEEDed:NTRansition <integer>
:STATus:QUEStionable:CALibration:EXTended:NEEDed:NTRansition?
Example STAT:QUES:CAL:EXT:NEED:NTR 2 Align EMI conducted is not required.
Preset 0
Min 0
Max 32767
Status Bits/OPC
dependencies
Sequential command
Initial S/W Revision Prior to A.02.00
124 EMI Receiver Mode Reference