8 Frequency Scan Measurement
Meas Setup
By Time
Specifies that the elements in the signal list will be sorted based on time.
Key Path
Meas Setup, Signal List, Sort Signals
Example CALC:FSC:SLIS:SORT:TYPE TIME
Initial S/W Revision A.07.00
Sort Order
Chooses between ascending and descending sort order.
Key Path
Meas Setup, Signal List, Sort Signals
Remote Command
:CALCulate:FSCan:SLISt:SORT:ORDer ASCending |DESCending
:CALCulate:FSCan:SLISt:SORT:ORDer?
Example CALC:FSC:SLIS:SORT:ORD DESC
CALC:FSC:SLIS:SORT:ORD
Dependencies This key is grayed out when a scan sequence is running. If the grayed out key is pressed, an advisory
message is generated. If the equivalent SCPI command is sent, this same message is generated as
part of a “–221, Settings conflict” warning.
Couplings Selecting the sort order (pressing the key or sending the equivalent SCPI command) will turn on the
Auto Sort state, even if that sort order was already selected.
Preset ASCending
State Saved Saved in instrument state.
Initial S/W Revision A.07.00
Auto Sort
If this is turned on, the measurement will perform signal sorting based on the sorting order and sorting type
selected before each new signal is added. Otherwise, the signal will be added into the signal list in
descending order of the trace amplitude level.
Key Path
Meas Setup, Signal List, Sort Signals
Remote Command
:CALCulate:FSCan:SLISt:SORT:AUTO ON|OFF|1|0
:CALCulate:FSCan:SLISt:SORT:AUTO?
Example CALC:FSC:SLIS:SORT:AUTO 1 Turn on auto sorting
CALC:FSC:SLIS:SORT:AUTO?
Notes If Scan Sequence is set to Remeasure, the signal in the list will only be auto sorted when a
Remeasure is completed.
Dependencies This key is grayed out when a scan sequence is running. If the grayed out key is pressed, an advisory
message is generated. If the equivalent SCPI command is sent, this same message is generated as
part of a “–221, Settings conflict” warning.