8 Frequency Scan Measurement
Sweep/ Control
Remote Command
:INITiate:PAUSe
Example
INIT:PAUS
Couplings When paused, the label on the key changes to “Resume”.
Backwards
Compatibility SCPI
:HOLD
Backwards Compatibility
SCPI Notes
The command above is included for ESU compatibility.
Initial S/W Revision A.07.00
Key Path
Sweep/ Control
Remote Command
:INITiate:RESume
Example
INIT:RES
Couplings When resumed, the label on the key changes to “Pause”.
Backwards
Compatibility SCPI
:INITiate[1]|2:CONMeas.
Backwards Compatibility
SCPI Notes
This command is included for ESU compatibility. The suffix [1]|2 is irrelevant to the measurement,
they are added solely for remote language compatibility.
Initial S/W Revision A.07.00
Start/ Stop
Allows you to start the scan, search or final measurement depending on the scan sequence selected.
When started, the label on the key changes to Stop. The meters display would turn off during scanning and
a final measurement. Pressing Stop will stop the current scan sequence and meters detector will be
started automatically.
Key Path
Softkey: Sweep/ControlHardkey: Restart
Remote Command
INITiate2:IMMediate
INITiate2:RESTart
Example
INIT2:IMMINIT2:REST
Notes This command is remote language compatible with ESU.
Couplings When started, the label on the key changes to Stop.
Initial S/W Revision A.07.00
Key Path
Sweep/ Control
Remote Command
[:SENSe]:FSCan:SCAN:ABORt
Example FSC:SCAN:ABOR
574 EMI Receiver Mode Reference