12 Disturbance Analyzer Measurement
Meas Setup
CALCulate:DANalyzer:DLISt:SORT:DAMPlitude ?
Example CALC:DAN:DLIS:SORT:DAMP PEAK
CALC:DAN:DLIS:SORT:DAMP?
Dependencies Selecting the detector amplitude type to be sorted (pressing the key or sending the equivalent SCPI
command) will turn on the Auto Sort state, even if that detector amplitude type was already
selected.
Preset PEAK
State Saved Saved in instrument state
Range Peak | Quasi Peak
Initial S/W Revision A.14.00
By Type of Disturbance
Specifies the disturbance list will be sorted based on type of disturbance.
Key Path
Meas Setup, Disturbance List, Sort Disturbances
Example CALC:DAN:DLIS:SORT:TYPE TDIS
Initial S/W Revision A.14.00
By Time
Specifies the disturbance list will be sorted based on time.
Key Path
Meas Setup, Disturbance List, Sort Disturbances
Example CALC:DAN:DLIS:SORT:TYPE TIME
Initial S/W Revision A.14.00
Sort Order
Chooses the sorting order.
Key Path
Meas Setup, Disturbance List, Sort Disturbances
Remote Command
CALCulate:DANalyzer:DLISt:SORT:ORDer ASCending |DESCending
CALCulate:DANalyzer:DLISt:SORT:ORDer?
Example CALC:DAN:DLIS:SORT:ORD DESC
CALC:DAN:DLIS:SORT:ORD?
Dependencies Selecting the sort order (pressing the key or sending the equivalent SCPI command) will turn on the
Auto Sort state, even if that sort order was already selected.
Preset ASCending
State Saved Saved in instrument state
Initial S/W Revision A.14.00