8 Frequency Scan Measurement
Meas Setup
Key Path
Meas Setup, Signal List
Initial S/W Revision A.07.00
Select Signal
Specifies the selected signal. The term “Selected signal” is used throughout this document to specify
which signal will be affected by the functions.
Key Path
Meas Setup, Signal List, Delete Signals
Notes Front panel only.
Dependencies Yes
Couplings When the signal list is cleared, this value will reset to 0.
Preset If the signal list is empty, =0
If the signal list is not empty, = number of peaks in signal list table
State Saved Saved in instrument state.
Min 0
Max Number of peaks in signal list table.
Initial S/W Revision A.07.00
Delete Signal
Deletes the selected signal. If you accessed this function from the front panel, the following message will
prompt you for confirmation to delete the selected signal from the signal list.
This will delete the selected signal. Are you sure you want to do this?
Please press Enter or OK to proceed. Press ESC or Cancel to close this dialog.
Key Path
Meas Setup, Signal List, Delete Signals
Remote Command
:CALCulate:FSCan:SLISt:DELete:SIGNal <integer>
Example CALC:FSC:SLIS:DEL:SIGN 4 Delete signal #4.
Notes If the signal list contains no data, an error is generated and this function cannot be performed.
When signals in the signal list were successfully deleted, an advisory is generated.
If the selected signal is not in the list, an error is generated and this function cannot be performed.
Dependencies This key is grayed out when a scan sequence is running. If the grayed out key is pressed, an advisory
message is generated. If the equivalent SCPI command is sent, this same message is generated as
part of a “–221, Settings conflict” warning.
Initial S/W Revision A.07.00
EMI Receiver Mode Reference 445