5 Mode Functions
Mode
Preset Not affected by a Preset
State Saved Not saved in state per se, the value will be the selected application when a Save is invoked.
Initial S/W Revision Prior to A.02.00
Application Identification Catalog (Remote Commands Only)
A catalog of the installed and licensed applications (Modes) can be queried for their identification.
"Application Catalog Number of Entries" on page 231
"Application Catalog Model Numbers" on page 231
"Application Catalog Revision" on page 231
"Application Catalog Options" on page 232
Application Catalog Number of Entries
Returns the number of installed and licensed applications (Modes).
Remote Command
:SYSTem:APPLication:CATalog[:NAME]:COUNt?
Example :SYST:APPL:CAT:COUN?
Preset Not affected by Preset
State Saved Not saved in instrument state.
Initial S/W Revision Prior to A.02.00
Application Catalog Model Numbers
Returns a list of Model Numbers for the installed and licensed applications (Modes).
Remote Command
:SYSTem:APPLication:CATalog[:NAME]?
Example :SYST:APPL:CAT?
Notes Returned value is a quoted string of a comma separated list of Model Numbers. Example, if SAMS
and Phase Noise are installed and licensed:
"N9060A,N9068A"
String length is COUNt * 7 – 1. (7 = Model Number length + 1 for comma. –1 = no comma for the
1st entry.)
Preset Not affected by a Preset
State Saved Not saved in instrument state.
Initial S/W Revision Prior to A.02.00
Application Catalog Revision
Returns the Revision of the provided Model Number.
EMI Receiver Mode Reference 231