12 Disturbance Analyzer Measurement
Meas Setup
DAN:CLIC:CORR?
Preset 0
State Saved Saved in instrument state
Min –100 dB
Max 100 dB
Initial S/W Revision A.14.00
Click Count
When turned on, specifies the Click Count value for the test to stop when the measured clicks is equal or
more than the entered clicks limit number.
Key Path
Meas Setup, Frequency
Remote Command
[:SENSe]:DANalyzer:CLICk:COUNt <integer>
[:SENSe]:DANalyzer:CLICk:COUNt?
[:SENSe]:DANalyzer:CLICk:COUNt:STATe ON|OFF|1|0
[:SENSe]:DANalyzer:CLICk:COUNt:STATe?
Example DAN:CLIC:COUN 10
DAN:CLIC:COUN?
DAN:CLIC:COUNt:STAT OFF
Preset 40
OFF
State Saved Saved in instrument state
Min 1
Max 1000
Initial S/W Revision A.14.00
By Δ Detector Type
Specifies that the elements in the signal list will be sorted based on the detector to limit delta type.
Key Path
Meas Setup, Signal List, Sort Signals, By ∆ Detector
Remote Command
:CALCulate:FSCan:SLISt:SORT:DLDelta DET1|DET2|DET3
:CALCulate:FSCan:SLISt:SORT:DLDelta?
Example CALC:FSC:SLIS:SORT:DLD DET2
CALC:FSC:SLIS:SORT:DLD?
Dependencies This key is grayed out when a scan sequence is running. If the grayed out key is pressed, an advisory
message is generated. If the equivalent SCPI command is sent, this same message is generated as
part of a “–221, Settings conflict” warning.
1070 EMI Receiver Mode Reference