6 System Functions
System
No confirmation is required when Alert is configured through a remote command.
Key Path
System, Alignments, Auto Align, Alert
Mode All
Example :CAL:AUTO:ALER DAY
Readback Text 24 hours
Status Bits/OPC
dependencies
Bit 14 is set in the Status Questionable Calibration register.
Initial S/W Revision Prior to A.02.00
7 days
With Auto Align Alert is set to 7 days the instrument will signal an alert after a time span of 168 hours since
the last successful full alignment (for example, Align Now, All or completion of a full Auto Align). You may
choose this selection in an environment where the temperature is stable on a weekly basis, at a modest
risk of accuracy degradations in excess of warranted performance. The alert is the Error Condition message
Align Now, All required”.
For front panel operation, confirmation is required for the customer to transition into this setting of Alert.
The confirmation dialog is:
No confirmation is required when Alert is configured through a remote command.
EMI Receiver Mode Reference 295

Table of Contents

Other manuals for Keysight PXA X Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight PXA X Series and is the answer not in the manual?

Keysight PXA X Series Specifications

General IconGeneral
BrandKeysight
ModelPXA X Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals