Instructions for use Zeus Infinity Empowered SW 2.n 351
Configuration
D Vol. agent defaults:
Specify the start values for the volatile anes-
thetic agents
If an anesthetic gas concentration control is se-
lected for the first time, the values in the respec-
tive control are displayed in the MixGas set-
tings dialog, see page 115.
Factory settings:
Calculation settings
The page More settings > Calculat. settings
opens.
Configuration of the calculation criteria for prelimi-
nary calculation of the VT and RR settings.
A Selection: Radford, Define, Off
Radford: The preliminary calculation of VT and
RR is based on the Radford algorithm. The
waveforms from Radf.VT (male), Radf.VT
(fem.) and Radf. RR are displayed in the
graphic.
Define: The preliminary calculation of VT and
RR is based on the markers entered under (E).
The waveforms from VT and RR are displayed
in the graphic.
Off: No preliminary calculation of VT and RR
takes place.
B Reset to Radford: Reset
The Reset button resets the user's definitions to
the default values.