11 APD Measurement
Meas Setup
Key Path
Meas Setup, Limits, APD Method
Example CALC:APD:LLIN:METH DLEV
State Saved Saved in instrument state.
Initial S/W Revision A.13.00
Probability of Time (Constant)
Defines the probability where the Disturbance Level limit is applies to.
Key Path
Meas Setup, Limits, APD Method, Disturbance Level
Remote Command
:CALCulate:APD:LLINe[1]:METHod:DLEVel:PROBability N1 | N2 | N3 | N4 |
N5 | N6 | N7
:CALCulate:APD:LLINe[1]:METHod:DLEVel:PROBability?
Example CALC:APD:LLIN:METH:DLEV:PROB N2 Specified limit at probability of 10 –2.
CALC:APD:LLIN:METH:DLEV:PROB?
Preset N1
State Saved Saved in instrument state.
Range N1|N2|N3|N4|N5|N6|N7
Initial S/W Revision A.13.00
N1
Sets the Probability of Time constant to 10–1.
Key Path
Meas Setup, Limits, APD Method, Disturbance Level, Probability of Time
Example CALC:APD:LLIN:METH:DLEV:PROB N1
Readback 101
Initial S/W Revision A.13.00
N2
Sets the Probability of Time constant to 10–2.
Key Path
Meas Setup, Limits, APD Method, Disturbance Level, Probability of Time
Example CALC:APD:LLIN:METH:DLEV:PROB N2
Readback 102
Initial S/W Revision A.13.00
946 EMI Receiver Mode Reference