EasyManua.ls Logo

Hach sc200 - Specifications of RS232 to RS485 Isolated Universal Controller

Hach sc200
875 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Specifications
Table 2-1.
Specifications of RS232 to RS485 Isolated Universal Controller
POWER SUPPLY Input Voltage 24 VDC
Permissible Range 20.4 to 28.8 VDC
Maximum Input Current 35 mA @ 24 VDC
RS232 Input voltage ±20 VDC maximum
Cable Length 15 m maximum (50 feet)
Interface Connectors RJ11 or screw-type, 2.5 mm, 2 terminal
RS422/RS485 Input Voltage -7 to +12V differential maximum
Cable Type Shielded twisted pair,
in compliance with EIA RS485
Cable Length 1200m maximum (4000 feet)
Interface Connector Screw-type, 2.5 mm, 2 terminal
GENERAL Isolation yes
Size 110.8 (L) x 22.5 (W) x 75 (H) mm
[4.38 (L) x 0.89 (W) x 2.96 (H)"]
Dip Switch Setting
S1 S2
12 3 4 1 2 34
OFF OFF OFF OFF OFF ON OFF ON
Figure 2-10. Dip Switch Settings
2-10

Table of Contents