EasyManua.ls Logo

Hach sc200 - Procedure 6. Setting the Current Time; Procedure 7. Setting the Current Date; To Update the Ultima Gas Monitor Internal Clock

Hach sc200
875 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
9. Aim the Controller at the sensor.
a. Press the 0 button to enable the alarm
as a non-energized alarm.
b. Press the 1 button to enable the alarm as an energized alarm.
The Ultima X Series Gas Monitor shows the alarm setpoint
value and the status:
LATCH/UNLATCH, INCR/DECR, ENER/DENERG.
Procedure 6. Setting the Current Time
The Ultima Gas Monitor is factory-set to Eastern Standard Time.
To change it in the Controller and in the Ultima Gas Monitor:
1. Press and hold the TIME button.
The display prompts: ##:##
2. Using the number keys, enter the correct time in the military format
(e.g., 4:00 P.M. = 16:00). (Leading zeros required.)
3. Press the ENTER button.
To update the Ultima Gas Monitor internal clock:
1. Press the SEND button.
2. Aim the Controller at the sensor and press the TIME button.
NOTE: Time and date are updated with this command.
The Ultima Gas Monitor displays the current time and date
for five seconds.
Procedure 7. Setting the Current Date
The Ultima Gas Monitor is factory-set to the current date.
To change it in the Controller and in the Ultima Gas Monitor:
1. Press and hold the DATE button to enter the correct date.
The display prompts: MM-DD-YYYY
2. Use number keys to enter correct date (leading zeros required):
MM = Months
DD = Date
• YYYY = Year
3. Press the ENTER button.
3-11