Procedures (see TABLE 3-1)
Procedure 1. Setting the Average Time Interval
The average, minimum and maximum gas concentration values are
gathered over the last time interval set by the Controller. This procedure
changes the time interval used for the average, minimum, and
maximum gas concentration value calculations of the Sensor.
To Change the Average Time Interval
1. Press the SEND button.
The display prompts: SEND?
2. Press the AVG button.
The display prompts: Set Avg Term +
3. Press the + or - button to scroll through the available list:
Every 1 Hour
Every 8 Hours
Every 24 Hours.
4. When the desired time interval displays, aim the controller at the
sensor and press the ENTER button.
The Monitor will show the time interval selected
for five seconds
The Monitor time interval is now set to the desired selection.
The Monitor displays the average, minimum, or maximum gas
concentration reading over the selected interval. This reading
is updated at the end of the selected average interval.
Procedure 2. Setting the Span Value
Monitors are shipped with TABLE 2-1 default span gas values. MSA
calibration cylinders are available for most of these pre-set span gas
concentrations; if an alternate span gas value is needed, the Monitor
span calibration value must be changed.
To Change the Calibration Span Gas Value of the Ultima Sensor
1. Press the SEND button.
The display prompts: SEND?
2. Press the SPAN button.
The display prompts: Span Gas Type +
3-3

Table of Contents

Other manuals for Hach sc200

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hach sc200 and is the answer not in the manual?

Hach sc200 Specifications

General IconGeneral
Enclosure RatingNEMA 4X/IP66
Operating Temperature-20 to 60 °C (-4 to 140 °F)
Ambient Temperature-20 to 60 °C (-4 to 140 °F)
Enclosure MaterialPolycarbonate
Operating Humidity0 to 95% non-condensing
Relay OutputsUp to 4 SPDT relays
Storage Conditions-40 to 70 °C (-40 to 158 °F)
DisplayGraphical LCD
Inputs2 digital sensor inputs
Outputs4-20 mA
Power Supply100 to 240 VAC, 50/60 Hz
Analog Outputs2 or 4, 0/4-20 mA
CertificationsCE
CommunicationModbus RTU, HART
Dimensions144 x 144 mm
MountingPanel mount
Output4-20 mA
Power Requirements (Voltage)100 to 240 VAC
Warranty1 year
Weight1.2 kg

Related product manuals