Table 3-18. Information Flags Word #3
(Read at address Base+26)
NAME BITS FUNCTION DESCRIPTION
Overrange Flag sensor #1 0 Set if TRUE
Overrange Flag sensor #21 1 Set if TRUE
Overrange Flag sensor #31 2 Set if TRUE
LOC Flag sensor #1 1 3 Set if TRUE
LOC Flag sensor #2 1 4 Set if TRUE
LOC Flag sensor #3 1 5 Set if TRUE
Parameter Fault Sensor #1 6 Set if TRUE
Parameter Fault Sensor #2 7 Set if TRUE
Parameter Fault Sensor #3 8 Set if TRUE
Warm up Sensor #1 1 9 Set if TRUE
Warm up Sensor #2 1 10 Set if TRUE
Warm up Sensor #3 1 11 Set if TRUE
Sensor Configuration Reset #1 12 Set if TRUE
Sensor Configuration Reset #2 13 Set if TRUE
Sensor Configuration Reset #3 14 Set if TRUE
Not Used 15
Table 3-19. Information Flags Word #4
(Read at address Base+257)
NAME BITS FUNCTION DESCRIPTION
Underrange Average Interval S1 0 Set if TRUE
Underrange Average Interval S2 1 Set if TRUE
Underrange Average Interval S3 2 Set if TRUE
Overrange Average Interval S1 3 Set if TRUE
Overrange Average Interval S2 4 Set if TRUE
Overrange Average Interval S3 5 Set if TRUE
Sensor Warning Sensor #1 6 Set if TRUE
Sensor Warning Sensor #2 7 Set if TRUE
Sensor Warning Sensor #3 8 Set if TRUE
Not Used 9
Not Used 10
Not Used 11
Not Used 12
Not Used 13
Not Used 14
Not Used 15