Section 5 Operation
5.1 Sensor Setup
When a sensor is initially installed, the serial number of the sensor will be displayed as the
sensor name. To change the sensor name refer to the following instructions:
1. Select Main Menu.
2. From the Main Menu, select SENSOR SETUP and confirm.
3. Highlight the appropriate sensor if more than one sensor is attached and confirm.
4. Select CONFIGURE and confirm.
5. Select EDIT NAME and edit the name. Confirm or cancel to return to the Sensor
Setup menu.
5.2 Sensor Data Logging
The sc controller provides one data log and one event log for each sensor. The data log
stores the measurement data at selected intervals. The event log stores a variety of
events that occur on the devices such as configuration changes, alarms, warning
conditions, etc. The data log and the event log can be read out in a CSV format. For
downloading the logs please refer to the controller user manual.
5.3 Sensor Diagnostics Menu for pH and ORP Menu
SELECT SENSOR (if more than one sensor is attached)
STATUS
ERROR LIST See section 7.1 on page 37.
WARNING LIST See section 7.2 on page 37.
5.4 pH Sensor Setup Menu
SELECT SENSOR (if more than one sensor is attached)
CALIBRATE
1-POINT AUTO Calibration with a single buffer — normally pH 7.
2-POINT AUTO Calibration with two buffers — normally pH 7 and pH 4 or 10.
1-POINT MANUAL Calibration against a single known sample.
2-POINT MANUAL Calibration against two samples, both with a known pH.
TEMP ADJUST Adjust the displayed temperature by up to ± 15 °C.
DEFAULT SETUP Restores the system to the original factory calibration.