15
Installation
shown in section 3.1.1 on page 11. The non-sensor end is wired to the sc100 controller in
a hazardous location as shown in section 3.1.2 on page 13.
3.3.1 Wiring the Digital Gateway
DANGER
The sc100 and certain versions of the sensor are suitable for use in Class 1,
Division 2, Groups A, B, C, D Hazardous Locations . See Control Drawing 58600-78
in the sc100 Controller Manual, Cat. No. 58600-18 for acceptable sensor versions
and installation requirements.
DANGER
Le sc100 et certaines versions du capteur peuvent être utilisés dans des endroits
dangereux de la Classe 1, Division 2, Groupes A, B, C, D. Reportez-vous au schéma
de contrôle 58600-78 du Manuel du contrôleur sc100, Réf. 58600-18 pour connaître
les versions des capteurs admises et les conditions d'installation.
DANGER
Explosion hazard. Do not connect or disconnect equipment unless power has been
switched off or the area is known to be non-hazardous.
DANGER
Risque d’explosion. Couper le courant ou s’assurer que l’emplacement est designe
non dangereux avant de replacer le aucon composant.
1. Route the cable from the sensor through the strain relief in the digital gateway then
properly terminate the wire ends (see Figure 8).
Note: Do not tighten the strain relief until the digital gateway is wired and the two halves are
threaded securely together.
2. Insert the wires as shown in Table 4 and Figure 9.
3. Make sure the O-ring is properly installed between the two halves of the digital
gateway and thread the two halves together. Hand tighten.
4. Tighten the strain relief to secure the sensor cable.
5. Connect the digital gateway to the controller.
sc100 Non-Hazardous Location Instructions—section 3.1.1 on page 11.
sc100 Hazardous Location Instructions—section 3.1.2 on page 13g
sc1000 Connection Instructions—Refer to section 3.2 on page 14.
Figure 8 Proper Wire Preparation and Insertion
1. Strip ¼-inch of insulation. 1. Seat insulation against connector with no bare wire exposed.