F1-Add Units to Network
• Password protected; see FIGURE 6-3
• Password is two levels:
• Level 1 is factory default
(contact an MSA Service Representative if needed)
• Level 2 is user-configurable (Default = 0).
• (FIGURE 6-4) Use this screen to:
• initialize a new Ultima X
3
sensor or
• read all units on the network.
F14- Select F14 once the following three steps are completed:
1) all Ultima X
3
sensors (up to 25 X
3
transmitters with up to three
sensors each) are wired to the network and have unique IDs.
2) network IDs are set by the Ultima Controller (P/N 809086)
(Valid IDs are 2 - 26).
NOTE: ID 0 is the broadcast modbus message; ID 1 is the
ModCon75 address.
3) Baud rate = 19200 and Parity = even.
• This process may take several minutes.
Figure 6-3. Password Entry
6-3