Section 2,
Typical Communication Cable Wiring
for Signal Boost (FIGURE 2-1)
Operation
Display:
Displays sensor type and gas level for each sensors
Cycles through the sensors
Sensor number graphic identifies sensor number
(see FIGURE 2-2)
Latches on an Alarm or Fault condition and requires user
acknowledgment prior to resuming cycling
If multiple conditions exist, reveals the subsequent alarm/fault
condition when one condition is acknowledged
Resumes cycling through display screens when all conditions are
acknowledged.
Alarm and fault conditions are indicated by alarm and the
corresponding number of the alarm level(s) that are activated
Relays:
Gas processing and alarm activation continues even when a
display is locked on a screen
Are common to all sensors by default
Can be assigned specifically to a sensor by enabling only one level
of alarm for each sensor
This may be accomplished through an Ultima X Controller
(see Section 2, "Calibration") consistent with the existing
Ultima X procedures or through a ModBUS controller
Form C contacts, 5 A resistive, 30 VDC, 250 VAC
May be configured as
Normally energized/de-energized
Increasing/decreasing level alarm
Latching/non-latching
2-1