Table 3-16. Information Flags Word #1 –
(Read at address Base+254)
NAME BITS FUNCTION DESCRIPTION
Sen
sor # 1 Disabled 0 0 = enabled, 1 = disabled
Sensor # 2 Disabled 1 0 = enabled, 1 = disabled
Sensor # 3 Disabled 2 0 = enabled, 1 = disabled
Alarm # 1 Sensor # 1 3 0 = clear, 1 = set
Alarm # 2 Sensor # 1 4 0 = clear, 1 = set
Alarm # 3 Sensor # 1 5 0 = clear, 1 = set
Alarm # 1 Sensor # 2 6 0 = clear, 1 = set
Alarm # 2 Sensor # 2 7 0 = clear, 1 = set
Alarm # 3 Sensor # 2 8 0 = clear, 1 = set
Alarm # 1 Sensor # 3 9 0 = clear, 1 = set
Alarm # 2 Sensor # 3 10 0 = clear, 1 = set
Alarm # 3 Sensor # 3 11 0 = clear, 1 = set
Cal Fault Condition Sensor #1 12 0 = Zero, 1 = Span
Cal Fault Condition Sensor #2 13 0 = Zero, 1 = Span
Cal Fault Condition Sensor #3 14 0 = Zero, 1 = Span
Not Used 15
Table 3-17. Information Flags Word #2
(Read at address Base+255)
NAME BITS FUNCTION DESCRIPTION
Configuration Reset 0 Set if TRUE
Fault RAM Main 1 Set if TRUE
FAULT FLASH MAIN 2 Set if TRUE
EEPROM WRITE ERROR 3 Set if TRUE
MUX FAULT 4 Set if TRUE
FAULT INCOMPATIBLE
SENSOR #1 5 Set if TRUE
FAULT INCOMPATIBLE
SENSOR #2 6 Set if TRUE
FAULT INCOMPATIBLE
SENSOR #3 7 Set if TRUE
Quick under range sensor #1 8 Set if TRUE
Quick under range sensor #2 9 Set if TRUE
Quick under range sensor #3 10 Set if TRUE
Under range Sensor #1 11 Set if TRUE
Under range sensor #2 12 Set if TRUE
Under range sensor #3 13 Set if TRUE
Atex Enabled 14 0 = Disabled, 1 = Enabled
Swap Delay 15 0 = Disabled, 1 = Enabled
3-12