4. Press the ENTER button.
The display prompts: CalHour HH.
5. Enter the two-digit hour desired
(from 0 to 23 hours YPM = 16 hours). (Leading zero is required.)
6. Aim the controller at the sensor and press the ENTER button.
The Ultima Gas Monitor displays the hour selected
The Ultima Gas Monitor is now programmed to auto-calibrate
at the desired time.
Procedure 10. Setting the Date of the
Next Scheduled Ultima Gas Monitor Calibration
To disable the auto-calibration of the Ultima Gas Monitor, set the date of
the next calibration to 12/31/94.
1. Press the SEND button.
The display prompts: SEND?
2. Press the CAL button.
The display prompts: Sel Cal Action +
.
3. Press either the + or - button and scroll to the NextCal Date
display message.
4. Press the ENTER button.
The display prompts: MM-DD YYYY.
5. Enter a valid eight-digit date (month-day-year).
6. Aim the controller at the sensor and press the ENTER button.
(Leading zeros are required.)
The Ultima Gas Monitor displays the future date
auto-calibration will start
The Ultima Gas Monitor is now programmed to auto-calibrate
on the desired date.
Procedure 11. Changing the MUX Address
1. Viewing the current Mux Address, aim the Controller at the Ultima
Gas Monitor.
2. Press the ADDRESS button.
The current address of the sensor displays.
NOTE: Additional pressing of the ADDRESS button increments
the address.
3-14

Table of Contents

Other manuals for Hach sc200

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hach sc200 and is the answer not in the manual?

Hach sc200 Specifications

General IconGeneral
Enclosure RatingNEMA 4X/IP66
Operating Temperature-20 to 60 °C (-4 to 140 °F)
Ambient Temperature-20 to 60 °C (-4 to 140 °F)
Enclosure MaterialPolycarbonate
Operating Humidity0 to 95% non-condensing
Relay OutputsUp to 4 SPDT relays
Storage Conditions-40 to 70 °C (-40 to 158 °F)
DisplayGraphical LCD
Inputs2 digital sensor inputs
Outputs4-20 mA
Power Supply100 to 240 VAC, 50/60 Hz
Analog Outputs2 or 4, 0/4-20 mA
CertificationsCE
CommunicationModbus RTU, HART
Dimensions144 x 144 mm
MountingPanel mount
Output4-20 mA
Power Requirements (Voltage)100 to 240 VAC
Warranty1 year
Weight1.2 kg