Appendix A,
Port 3 Ethernet Port /
OPC Server
This device has the capability to communicate to third party HMI
packages, such as Wonderware, Intellution, etc.
Communication is done through the ethernet board, which is
located on the ModCon75 Controller port 3 location.
Communication is ModBus over ethernet.
One can communicate through the OPC server (available as a
download on http:\\www.unitronics.com/download.htm) or through
an ethernet server provided by the third party vendor.
Using the OPC server, the following will apply:
Mapping of variables:
Name = Modcon75
IP Address = 192.168.1.11
Subnet Mask = 255.255.255.0
Gateway = 192.168.1.254
Retrieving gas value for specific sensor:
Value is split into two registries;
For example, ID#2 sensor #1:
MI300 provides value to the
left of the decimal point
MI325 provides value to the
right of the decimal point.
(See Appendix B, "Memory Integers").
When using the ethernet server (provided by the third party
vendor), the addressing may be off-set by one.
A-1