Section 4,
Maintenance
The Ultima Controller and Ultima Calibrator are rated as
intrinsically safe for Class I, Division 1, Groups A, B, C and D
areas during normal operation. All maintenance procedures
must be performed in a non-hazardous area. Failure to
comply with this warning can result in serious personal
injury or death.
Batteries
The Ultima Controller is approved for use with two Duracell
MN1500 "AA" Alkaline cells.
The Ultima Calibrator is approved for use with either two Duracell
MN2400 or two Eveready E92 "AAA" Alkaline cells.
Determining a Low Battery Condition
on the Ultima Controller
To Determine if Battery Replacement is Necessary
1. Turn the controller unit ON.
After the unit completes its initialization,
the display shows ID CODE, READY or ready.
2. If the display shows the ID CODE message, enter the correct code
(see "Using the ID CODE Feature") to reach the ready prompt.
3. The READY display is the low battery indicator:
• If READY appears (upper case letters), the batteries are OK
If ready appears (lower case letters), the batteries are weak
and must be replaced.
Replacing the Batteries
To Install New Batteries in the Ultima Controller
1. Turn OFF the Ultima Controller.
2. Remove the four screws from the rear of the case and carefully
pull the rear cover away from the unit.
3. Remove the two weak batteries from the battery holder and
dispose of properly.
"
WARNING
4-1

Table of Contents

Other manuals for Hach sc200

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hach sc200 and is the answer not in the manual?

Hach sc200 Specifications

General IconGeneral
Enclosure RatingNEMA 4X/IP66
Operating Temperature-20 to 60 °C (-4 to 140 °F)
Ambient Temperature-20 to 60 °C (-4 to 140 °F)
Enclosure MaterialPolycarbonate
Operating Humidity0 to 95% non-condensing
Relay OutputsUp to 4 SPDT relays
Storage Conditions-40 to 70 °C (-40 to 158 °F)
DisplayGraphical LCD
Inputs2 digital sensor inputs
Outputs4-20 mA
Power Supply100 to 240 VAC, 50/60 Hz
Analog Outputs2 or 4, 0/4-20 mA
CertificationsCE
CommunicationModbus RTU, HART
Dimensions144 x 144 mm
MountingPanel mount
Output4-20 mA
Power Requirements (Voltage)100 to 240 VAC
Warranty1 year
Weight1.2 kg