8. To Enable the chosen setpoint, press the 1 button
The display prompts: Latched 0=N 1=Y.
Aim the Controller at the sensor:
a. Press the 0 button to enable the alarm in unlatched mode
or
b. Press the 1 button to enable the alarm in latched mode.
The Ultima Gas Monitor will show both the status (on U if
unlatched or on L if latched) and the alarm setpoint value.
Enabling/Disabling and setting the Mode of the
Three Ultima X Series Alarm Setpoints
1. Press the SEND button.
The display prompts: SEND?.
2. Press the ALARM button.
The display prompts: 0=SetPt 1=OnOFF.
3. Press the 1 button.
The display prompts: Alm Set Point #.
4. Using the NUMBER buttons, enter the desired alarm setpoint 1, 2,
or 3; then, press the ENTER button. (Invalid entries are ignored.)
• The DEL button can be used to delete number entries before
the ENTER button is pressed.
The display prompts: AlmSPnt 1=E 0=D.
5. To Disable the chosen setpoint, aim the Controller at the sensor
and press the 0 button.
The Ultima X Series Gas Monitor will show both the status
(OFF) and the alarm setpoint value.
6. To enable the chosen setpoint, press the 1 button.
The display prompts: Latched 0=N 1=Y.
7. To set the alarm as a latching alarm, press the 1 button;
to set the alarm as unlatching, press the 0 button.
(Invalid entries are ignored.)
The display prompts: 0=DOWN 1=UP.
8. To set the alarm as a downward acting alarm, press the 0 button;
to set the alarm as an upward acting alarm, press the 1 button.
(Invalid entries will be ignored.)
The display prompts: 0=NONEN 1=ENGZD.
3-10

Table of Contents

Other manuals for Hach sc200

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hach sc200 and is the answer not in the manual?

Hach sc200 Specifications

General IconGeneral
Enclosure RatingNEMA 4X/IP66
Operating Temperature-20 to 60 °C (-4 to 140 °F)
Ambient Temperature-20 to 60 °C (-4 to 140 °F)
Enclosure MaterialPolycarbonate
Operating Humidity0 to 95% non-condensing
Relay OutputsUp to 4 SPDT relays
Storage Conditions-40 to 70 °C (-40 to 158 °F)
DisplayGraphical LCD
Inputs2 digital sensor inputs
Outputs4-20 mA
Power Supply100 to 240 VAC, 50/60 Hz
Analog Outputs2 or 4, 0/4-20 mA
CertificationsCE
CommunicationModbus RTU, HART
Dimensions144 x 144 mm
MountingPanel mount
Output4-20 mA
Power Requirements (Voltage)100 to 240 VAC
Warranty1 year
Weight1.2 kg