Table 3-11. Control Words (Read/Write)
DESCRIPTION ADDRESS POSSIBLE VALUES
Command Word 1 Base+301 0 to 32767, See TABLE 3-12
Command Word 2 Base+302 0 to 32767, See TABLE 3-13
Table 3-12. ModBUS Command Word 1
(Read at address Base+301/Write Coils 1 through 16)
NAME BITS COIL FUNCTION DESCRIPTION
Start Full ICAL Calibration -
Sensor 1 0 1 Rtn's fault if any
Calibration in progress
Start Full ICAL Calibration -
Sensor 2 1 2 Rtn's fault if any
Calibration in progress
Start Full ICAL Calibration -
Sensor 3 2 3 Rtn's fault if any
Calibration in progress
Start Standard Full Calibration -
Sensor 1 3 4 Rtn's fault if any Calibration in progress
Start Standard Full Calibration -
Sensor 2 4 5 Rtn's fault if any Calibration in progress
Start Standard Full Calibration -
Sensor 3 5 6 Rtn's fault if any Calibration in progress
Start Standard Zero Calibration
Sensor 1 6 7 Rtn's fault if any Calibration in progress
Start Standard Zero Calibration
Sensor 2 7 8 Rtn's fault if any Calibration in progress
Start Standard Zero Calibration
Sensor 3 8 9 Rtn's fault if any Calibration in progress
Start UCAL Calibration -
Sensor 1 9 10 Rtn's fault if any Calibration in progress
Start UCAL Calibration -
Sensor 2 10 11 Rtn's fault if any Calibration in progress
Start UCAL Calibration -
Sensor 3 11 12 Rtn's fault if any Calibration in progress
Step UCAL 12 13 1 to step
Abort Calibration (any) 13 14 1 to abort
Reserved for future use 14 15
Not used 15 16