EasyManuals Logo

Hach sc200 User Manual

Hach sc200
875 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #92 background imageLoading...
Page #92 background image
38
Troubleshooting
7.3 Troubleshooting the pH Sensor
Clean the sensor using the procedure described in section 6.2 on page 34. If the
measuring system cannot be calibrated after cleaning, replace the standard cell solution
and salt bridge (see section 6.2.1 on page 35) and try calibrating again. If the measuring
system still cannot be calibrated, check the sensor operation.
Some simple tests using the sc100 or a multimeter and two pH buffers will determine if the
pH sensor is operating properly. The use of pH 7 and pH 4 buffers is preferred but pH 10
can be used in place of pH 4 if it more closely covers the measurement range of interest.
Determine if the sensor has integral digital electronics or uses an external digital gateway.
If the sensor uses a digital gateway, it will be hard-wired to the gateway through terminal
connections inside the digital gateway enclosure. If the sensor uses the digital gateway
and therefore does not have integral digital electronics, proceed with section 7.3.1. If the
sensor has integral digital electronics, move to section 7.3.2 on page 39.
7.3.1 Troubleshooting a pH Sensor without Integral Digital Electronics
1. Disconnect the red, green, yellow, and black sensor wires from the digital gateway.
2. Place the sensor in a pH 7 buffer. Before continuing, allow the temperatures of the
sensor and buffer to equalize to approximately 25 °C (70 °F).
3. Verify that the sensor temperature element (300 ohm thermistor) is operating properly
by measuring the resistance between the yellow and black wires. The reading should
be between 250 and 350 ohms at approximately 25 °C (70 °F).
4. Reconnect the yellow and black wires.
5. Connect the multimeter (+) lead to the red wire and (–) lead to the green wire. With the
sensor in the pH 7 buffer, measure the dc millivolts. The sensor offset reading should
be within the factory-specified limits of –50 and +50 mV. If it is, record the millivolt
value reading and continue with step 6. If the reading is outside these limits,
discontinue this test and contact Technical Support.
6. With the multimeter still connected, rinse the sensor with water and place it in either
pH 4 or pH 10 buffer. Allow the temperatures of the sensor and buffer to equalize to
approximately 25 °C (70 °F) then measure the sensor span reading as shown in Table
8 and Table 9 on page 39.
Span Reading in pH 4 Buffer
With the sensor in pH 4 buffer, the sensor span reading should be at least +160 mV more
than the offset reading taken in step 5.
Table 8 Typical Span Reading Examples (pH 4 buffer)
Offset Reading
(in pH 7 buffer)
Span Reading (in pH 4 buffer)
–50 mV +110 mV
–25 mV +135 mV
0 mV +160 mV
+25 mV +185 mV
+50 mV +210 mV

Table of Contents

Other manuals for Hach sc200

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hach sc200 and is the answer not in the manual?

Hach sc200 Specifications

General IconGeneral
Enclosure RatingNEMA 4X/IP66
Operating Temperature-20 to 60 °C (-4 to 140 °F)
Ambient Temperature-20 to 60 °C (-4 to 140 °F)
Enclosure MaterialPolycarbonate
Operating Humidity0 to 95% non-condensing
Relay OutputsUp to 4 SPDT relays
Storage Conditions-40 to 70 °C (-40 to 158 °F)
DisplayGraphical LCD
Inputs2 digital sensor inputs
Outputs4-20 mA
Power Supply100 to 240 VAC, 50/60 Hz
Analog Outputs2 or 4, 0/4-20 mA
CertificationsCE
CommunicationModbus RTU, HART
Dimensions144 x 144 mm
MountingPanel mount
Output4-20 mA
Power Requirements (Voltage)100 to 240 VAC
Warranty1 year
Weight1.2 kg

Related product manuals